T200U-NOy ANALYZER ADDENDUM (for use with T200 Operators Manual, PN 06858 and with the T200U Addendum, PN 06861) © TELEDYNE ADVANCED POLLUTION INS
Model T200EU-NOy Option Manual Addendum 1.0 Introduction 8 07303C DCN6646 This page intentionally left blank.
07303C DCN6646 9 2.0 SPECIFICATIONS 2.1 Specifications Table 1 Specifications for the External Converter and the 501Y Bypass Pump Converter Remo
Model T200EU-NOy Option Manual Addendum 2.0 Specifications 10 07303C DCN6646 This page intentionally left blank.
07303C DCN6646 11 3.0 GETTING STARTED The NOy converter has been designed to operate with the T200U low level NOx analyzer, with some additional m
Model T200U-NOy Option Manual Addendum 12 07303C DCN6646 3.2 Electrical and Pneumatic Connections NOTE To maintain compliance with EMC standards,
Model T200U-NOy Option Manual Addendum 07303C DCN6646 13 Figure 1. Rear Panels: T200U-NOy Analyzer (top) and 501Y (center); Converter (bottom)
Model T200U-NOy Option Manual Addendum 14 07303C DCN6646 Figure 2. Rear Panel Pneumatic Connections Note: Pneumatic tubes and rear panel labels in
Model T200U-NOy Option Manual Addendum 07303C DCN6646 15 3.3 Initial Operation 1. After confirming proper supply voltage, turn on the power to
Model T200U-NOy Option Manual Addendum 16 07303C DCN6646 This page intentionally left blank.
07303C DCN6646 17 4.0 THE T200U-NOY EXTERNAL CONVERTER 4.1 Principle of Operation The T-API Model T200U-NOy is designed to measure the concentr
Model T200U-NOy Option Manual Addendum 18 07303C DCN6646 This page intentionally left blank.
07303C DCN6646 19 5.0 CALIBRATION AND CAL CHECKS Unlike most NOx analyzers, the T200U-NOy does not have a sample inlet port on its rear panel. T
Model T200U-NOy Option Manual Addendum 20 07303C DCN6646 5.1 Calibration or Cal Check Procedure Calibration of the analyzer can be checked or adjus
Model T200U-NOy Option Manual Addendum 07303C DCN6646 21 Table 5 Span Calibration Procedure Step Action Comment 1. Press CAL The analyzer ent
Model T200U-NOy Option Manual Addendum 22 07303C DCN6646 This page intentionally left blank.
07303C DCN6646 23 6.0 MAINTENANCE 6.1 Maintenance Schedule The schedule for preventative maintenance is presented in Table 6 below. Qualified Mai
Model T200U-NOy Option Manual Addendum 24 07303C DCN6646 6.2 Replacing the Sample Particulate Filters The particulate filter should be inspected oft
Model T200U-NOy Option Manual Addendum 07303C DCN6646 25 Figure 3. Replacing the Particulate Filter
Model T200U-NOy Option Manual Addendum 26 07303C DCN6646 6.3 Checking Analyzer Flow Rate The external Pump is capable of maintaining the reaction c
Model T200U-NOy Option Manual Addendum 07303C DCN6646 27 8. Turn the 501Y power back on. The insulation can emit a burnt odor for the first 24 ho
07303C DCN6646 i SAFETY MESSAGES Important safety messages are provided throughout this manual for the purpose of avoiding personal injury or instru
Model T200U-NOy Option Manual Addendum 28 07303C DCN6646 Figure 5. Pneumatics
Model T200U-NOy Option Manual Addendum 07303C DCN6646 29 6.6 Checking for Leaks If a leak checker is not available, it is possible to leak check t
Model T200U-NOy Option Manual Addendum 30 07303C DCN6646 If you have a leak checker: 1. Turn off power to the analyzer and 501Y, and unplug the ext
07303C DCN6646 31 7.0 TROUBLESHOOTING, ADJUSTMENTS CAUTION! The operations outlined in this section must be performed by qualified maintenance per
Model T200U-NOy Option Manual Addendum 32 07303C DCN6646 7.1.3 501Y Bypass Pump Diagnostic Procedures The 501Y bypass pump is capable of maintainin
Model T200U-NOy Option Manual Addendum 07303C DCN6646 33 Indications that would require re-programming are: Failure / replacement of controller
Model T200U-NOy Option Manual Addendum 34 07303C DCN6646 SECONDARY MENU (The fixed characteristics of the system). PRESS DISPLAY ACTION SEL hold u
Model T200U-NOy Option Manual Addendum 07303C DCN6646 35 Figure 7 Electrical Diagram
Model T200U-NOy Option Manual Addendum 36 07303C DCN6646 This page intentionally left blank.
07303C DCN6646 37 8.0 T200U-NOY SPARE PARTS LIST The following spare parts list supplements the spare parts list for the T200U analyzer. Note U
ii 07303C DCN6646 CONSIGNES DE SÉCURITÉ Des consignes de sécurité importantes sont fournies tout au long du présent manuel dans le but d’éviter des
Model T200U-NOy Option Manual Addendum 38 07303C DCN6646 Reference: 06848 11/8/2010 1:13:10 PM PARTNUMBER DESCRIPTION 041800600
Model T200U-NOy Option Manual Addendum 07303C DCN6646 39 Reference: 06848 11/8/2010 1:13:10 PM PARTNUMBER DESCRIPTION CN0000073
Model T200U-NOy Option Manual Addendum 40 07303C DCN6646 This page intentionally left blank.
07303C DCN6646 iii WARRANTY WARRANTY POLICY (02024F) Teledyne Advanced Pollution Instrumentation (TAPI), a business unit of Teledyne Instruments, In
Model 200EU-NOy Option Manual Addendum iv 07303C DCN6646 This page intentionally left blank.
Model 200EU-NOy Option Manual Addendum 07303C DCN6646 v TABLE OF CONTENTS 1.0 INTRODUCTION...
Model 200EU-NOy Option Manual Addendum vi 07303C DCN6646 List of Figures Figure 1. Rear Panels: T200U-NOy Analyzer (top) and 501Y (center); Convert
07303C DCN6646 7 1.0 INTRODUCTION The T200U-NOy analyzer consists of an external Converter designed to support a NOx analyzer by converting multip