MODEL 452 PROCESS OZONE SENSOR ©Teledyne Instruments Advanced Pollution Instrumentation Division (T-API) 9480 Carroll Park Drive San Diego, CA 92121-
Teledyne API M452 Ozone Sensor Instruction Manual, 02852, Rev. E 1-1 1 Product Descriptions 1.1 M452 Process Ozone Sensor The Teledyne API M452 is a
Teledyne API M452 Ozone Sensor Instruction Manual, 02852, Rev. E 1-2 INTENTIONALLY BLANK
Teledyne API M452 Ozone Sensor Instruction Manual, 02852, Rev. E 2-1 2 Specifications Note: All specifications contained herein are subject to change
Teledyne API M452 Ozone Sensor Instruction Manual, 02852, Rev. E 2-2 2.2 Performance Specifications Accuracy: Repeatability: 1% of Full Scale. 1% of
Teledyne API M452 Ozone Sensor Instruction Manual, 02852, Rev. E 2-3 2.6 Pressure Drop Figure 2-2 below shows the approximate pressure drop from the
Teledyne API M452 Ozone Sensor Instruction Manual, 02852, Rev. E 2-4 2.7 Options The M452 is available with the following options: 2.7.1 AC Power Ada
Teledyne API M452 Ozone Sensor Instruction Manual, 02852, Rev. E 3-1 3 Theory of Operation The detection of ozone molecules is based on absorption of
Teledyne API M452 Ozone Sensor Instruction Manual, 02852, Rev. E 3-2 The intensity of light is converted into a voltage by the detector/preamp module
Teledyne API M452 Ozone Sensor Instruction Manual, 02852, Rev. E 4-1 4 Installation 4.1 Unpacking Upon receiving the M452 please verify that no appar
Teledyne API M452 Ozone Sensor Instruction Manual, 02852, Rev. E 4-2 Analog Out -Analog Out +Power Common+15 Volt Supply+15 Volt SupplyDigital In #1D
Teledyne API M452 Ozone Sensor Instruction Manual, 02852, Rev. E i SAFETY MESSAGES NOTE The M452 contains no user serviceable parts. CAUTION There
Teledyne API M452 Ozone Sensor Instruction Manual, 02852, Rev. E 4-3 4.3.2 Analog Output The analog output is a 0-5 volt signal representing the ozon
Teledyne API M452 Ozone Sensor Instruction Manual, 02852, Rev. E 4-4 4.3.4 Status Outputs The M452 has four digital status outputs for indicating err
Teledyne API M452 Ozone Sensor Instruction Manual, 02852, Rev. E 4-5 4.3.5 RS232/485 Interface The M452 features a bi-directional digital serial inte
Teledyne API M452 Ozone Sensor Instruction Manual, 02852, Rev. E 4-6 4.4 Gas Connections Gas connections to the M452 are made using ¼” compression tu
Teledyne API M452 Ozone Sensor Instruction Manual, 02852, Rev. E 4-7 4.5 Start-Up and Calibration Procedure 1. Verify that the proper electrical con
Teledyne API M452 Ozone Sensor Instruction Manual, 02852, Rev. E 5-1 5 Sensor and System Troubleshooting This chapter gives guidelines for diagnosing
Teledyne API M452 Ozone Sensor Instruction Manual, 02852, Rev. E 5-2 5.2 Status Outputs Table 5-1 below describes the function of the status outputs.
Teledyne API M452 Ozone Sensor Instruction Manual, 02852, Rev. E 5-3 5.4 Invalid Reading The normal state for the Invalid Reading output is OFF. If
Teledyne API M452 Ozone Sensor Instruction Manual, 02852, Rev. E 5-4 5.6 Cell Dirty The normal state for the Cell Dirty output is OFF. If this outpu
Teledyne API M452 Ozone Sensor Instruction Manual, 02852, Rev. E 5-5 INTENTIONALLY BLANK
Teledyne API M452 Ozone Sensor Instruction Manual, 02852, Rev. E iii TABLE OF CONTENTS SAFETY MESSAGES...
Teledyne API M452 Ozone Sensor Instruction Manual, 02852, Rev. E iv 5 SENSOR AND SYSTEM TROUBLESHOOTING ...5-1 5.1 STATUS LED’S...
Teledyne API M452 Ozone Sensor Instruction Manual, 02852, Rev. E v LIST OF FIGURES FIGURE 2-1: M452A HIGH PURITY OZONE SENSOR...
Teledyne API M452 Ozone Sensor Instruction Manual, 02852, Rev. E vii WARRANTY POLICY ADVANCED POLLUTION INSTRUMENTATION DIVISION 02024c Prior to ship
Teledyne API M452 Ozone Sensor Instruction Manual, 02852, Rev. E viii THE WARRANTIES SET FORTH IN THIS SECTION AND THE REMEDIES THEREFORE ARE EXCLUSI
Commentaires sur ces manuels