Teledyne Stud Sensor 452 Manuel d'utilisateur Page 19

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 30
  • Table des matières
  • DEPANNAGE
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 18
Teledyne API M452 Ozone Sensor Instruction Manual, 02852, Rev. E
4-2
Analog Out -
Analog Out +
Power Common
+15 Volt Supply
+15 Volt Supply
Digital In #1
Digital In #2
Digital Out #1
RS485-A
Power Common
RS485-B
Digital Out Common
Digital Out #4
Digital Out #3
Digital Out #2
Power Supply
Common
+15V
-+
Ozone Concentration
Meter
(0-5 VDC)
or Data Acqusition Device
Status Display
or
PLC
Zero Cal = Closed
Figure 4-1: Electrical Connections
4.3.1 Power Supply
The M452 requires a +15 VDC power source capable of supplying 1.0 A.
DC power can be connected through the male DB-15 connector or through
the coaxial power jack. The coaxial power jack is configured so that the
ground connection is on the outside (shield) and the +15V connection is on
the center pin.
If power is to be supplied through the DB-15 connector, the positive
terminal of the power supply should be connected to pins 4 and 5 on the
15-pin connector and the common terminal should be connected to pins 3
and 15.
If the optional AC Power Adapter is used to provide power, it should be
connected to the coaxial power connector adjacent to the DB-15 connector
on the top of the M452. Use only the approved AC Power Adapter
provided by Teledyne API.
Vue de la page 18
1 2 ... 14 15 16 17 18 19 20 21 22 23 24 ... 29 30

Commentaires sur ces manuels

Pas de commentaire