Teledyne API Model 465L O
3
Monitor Operation Manual
28 05509F DCN6409
4.1.2. Exhaust Connection
Connect a ¼” OD vent line to the “Exhaust” fitting on the rear panel. This line must be vented
outdoors or to an appropriate discharge system capable of handling ozone. Local regulations
regarding the discharge of ozone to the atmosphere must be observed. Note that the monitor
may be equipped with an internal ozone scrubber for removal of ozone from the exhaust stream.
For safety reasons, the exhaust port must still be vented appropriately as described above.
Figure 4-6: Rear Panel Electrical I/O Connections
Figure 4-7: NEMA Electrical I/O Connections