Teledyne API Model M201E NH
3
Analyzer Operator Manual Configurable Analog Output Overview
05206H DCN5910 29
3.3 SELECTING AUTO OR MANUAL ANALOG OUTPUT CALIBRATION
The analog outputs configured for voltage mode can be calibrated either
automatically or manually. In its default mode the instrument is configured for
automatic calibration. Note: manual calibration should only be used for the 0.1V
range or in cases where the outputs must be closely matched to the
characteristics of the recording device. Outputs configured for automatic
calibration can be calibrated as a group or individually.
To select auto-calibration for a particular channel press the following: