Teledyne T101 Manuel d'utilisateur

Naviguer en ligne ou télécharger Manuel d'utilisateur pour Mesure, test et contrôle Teledyne T101. Teledyne T101 User's Manual Manuel d'utilisatio

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 371
  • Table des matières
  • DEPANNAGE
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 0
OPERATION MANUAL
MODEL T101
UV FLUORESCENCE H
2
S ANALYZER
Also supports operation of:
Model T102 Analyzer
(when used in conjunction with T102 Addendum, PN 07267)
© TELEDYNE ADVANCED POLLUTION INSTRUMENTATION
9480 CARROLL PARK DRIVE
SAN DIEGO, CA 92121-5201
USA
Toll-free Phone:
800-324-5190
Phone:
858-657-9800
Fax:
858-657-9816
Email:
api-sales@teledyne.com
Website:
http://www.teledyne-api.com/
Copyright 2011 - 2012 07266B DCN6845
Teledyne Advanced Pollution Instrumentation 08 June 2012
Vue de la page 0
1 2 3 4 5 6 ... 370 371

Résumé du contenu

Page 1 - S ANALYZER

OPERATION MANUAL MODEL T101 UV FLUORESCENCE H2S ANALYZER Also supports operation of: Model T102 Analyzer (when used in conjunction with T102 A

Page 2

viii This page intentionally left blank. 07266B DCN6485

Page 3

Operating Instructions Teledyne API – T101 Operation Manual 98 4.6.5. ELECTRICAL TEST The electrical test function creates a current, which substi

Page 4

Teledyne API – T101 Operation Manual Operating Instructions 99 4.6.6. LAMP CALIBRATION An important factor in accurately determining H2S concentrati

Page 5 - SAFETY MESSAGES

Operating Instructions Teledyne API – T101 Operation Manual 100 4.6.7. PRESSURE CALIBRATION A sensor at the exit of the sample chamber continuousl

Page 6 - CONSIGNES DE SÉCURITÉ

Teledyne API – T101 Operation Manual Operating Instructions 101 4.6.8. FLOW CALIBRATION The flow calibration allows the user to adjust the values of

Page 7 - WARRANTY

Operating Instructions Teledyne API – T101 Operation Manual 102 4.6.9. TEST CHANNEL OUTPUT When activated, output channel A4 can be used to report

Page 8

Teledyne API – T101 Operation Manual Operating Instructions 103 Once a TEST function is selected, the instrument begins to report a signal on the A4

Page 9 - REVISION HISTORY

Operating Instructions Teledyne API – T101 Operation Manual 104 To edit the instrument’s ID code, press: SETUP X.X COMMUNICATIONS MENU

Page 10 - 07266B DCN6485

Teledyne API – T101 Operation Manual Operating Instructions 105 4.7.2. COM PORT DEFAULT SETTINGS As received from the factory, the analyzer is set u

Page 11 - TABLE OF CONTENTS

Operating Instructions Teledyne API – T101 Operation Manual 106 The signals from these two connectors are routed from the motherboard via a wiring

Page 12

Teledyne API – T101 Operation Manual Operating Instructions 107 4.7.4. RS-485 CONFIGURATION As delivered from the factory, COM2 is configured for R

Page 13

ix TABLE OF CONTENTS 1. INTRODUCTION ...

Page 14

Operating Instructions Teledyne API – T101 Operation Manual 108 will automatically be assigned an IP address by the DHCP server (Section 4.10.6.2)

Page 15

Teledyne API – T101 Operation Manual Operating Instructions 109 4.7.6.2. Manually Configuring the Ethernet with Static IP Addresses 1. Connect a c

Page 16 - LIST OF APPENDICES

Operating Instructions Teledyne API – T101 Operation Manual 110 ENTR accept new settings EXIT ignores new settings SETUP X.X PRIMARY

Page 17 - LIST OF FIGURES

Teledyne API – T101 Operation Manual Operating Instructions 111 Configure the INSTRUMENT IP, GATEWAY IP and SUBNET MASK addresses: Internet Confi

Page 18 - LIST OF TABLES

Operating Instructions Teledyne API – T101 Operation Manual 112 4.7.6.3. Changing the Analyzer’s HOSTNAME The HOSTNAME is the name by which the

Page 19

Teledyne API – T101 Operation Manual Operating Instructions 113 Table 4-18. Internet Configuration Touchscreen Button Functions BUTTON FUNCTION <

Page 20

Operating Instructions Teledyne API – T101 Operation Manual 114 4.7.7. USB CONFIGURATION The analyzer can be operated through a personal computer

Page 21 - 1. INTRODUCTION

Teledyne API – T101 Operation Manual Operating Instructions 115 5. In the instrument’s SETUP>MORE>COMM>COM2 menu, make the following setti

Page 22 - 1.2. OPTIONS

Operating Instructions Teledyne API – T101 Operation Manual 116 After connecting a USB cable between your PC and the instrument, ensure their baud

Page 23

Teledyne API – T101 Operation Manual Operating Instructions 117 Figure 4-9. Jumper and Cables for Multidrop Mode Note If you are adding an instru

Page 24

TABLE OF CONTENTS Teledyne API – T101 Operation Manual x 4.4.3. Setup – DAS: Data Acquisition ...

Page 25 - DESCRIPTION

Operating Instructions Teledyne API – T101 Operation Manual 118 Analyzer Analyzer Analyzer Last Analyzer Female DB9Male DB9 RS-232 COM2 RS-232

Page 26 - 2.2.2. EMC

Teledyne API – T101 Operation Manual Operating Instructions 119 4.7.9. MODBUS SET UP The following set of instructions assumes that the user is fami

Page 27 - 3. GETTING STARTED

Operating Instructions Teledyne API – T101 Operation Manual 120 Example Read/Write Definition window: Example Connection Setup window: Exampl

Page 28 - 3.2. INSTRUMENT LAYOUT

Teledyne API – T101 Operation Manual Operating Instructions 121 4.7.10. COM PORT COMMUNICATION MODES Each of the analyzer’s serial ports can be conf

Page 29

Operating Instructions Teledyne API – T101 Operation Manual 122 Press the following buttons to select a communication mode for a one of the COMM

Page 30

Teledyne API – T101 Operation Manual Operating Instructions 123 4.7.11. COM PORT BAUD RATE To select the baud rate of one of the COM Ports, press:

Page 31

Operating Instructions Teledyne API – T101 Operation Manual 124 4.7.12. COM PORT TESTING The serial ports can be tested for correct connection and

Page 32 - 3.2.2. REAR PANEL

Teledyne API – T101 Operation Manual Operating Instructions 125 the DAS (Section 4.8). The APICOM manual, which is included with the program, contai

Page 33

Operating Instructions Teledyne API – T101 Operation Manual 126 Table 4-21. DAS Data Channel Properties PROPERTY DESCRIPTION DEFAULT SETTING RANGE

Page 34

Teledyne API – T101 Operation Manual Operating Instructions 127 Table 4-22. DAS Data Parameter Functions FUNCTION EFFECT PARAMETER Instrument-speci

Page 35 - 3.3. ELECTRICAL CONNECTIONS

Teledyne API – T101 Operation Manual TABLE OF CONTENTS xi 4.8.2.9. Starting Date ...

Page 36 - ANALOG OUT

Operating Instructions Teledyne API – T101 Operation Manual 128 4.8.2. DEFAULT DAS CHANNELS A set of default Data Channels has been included in th

Page 37

Teledyne API – T101 Operation Manual Operating Instructions 129 LIST OF PARAMETERS NAME: PNUMTC EVENT: ATIMER REPORT PERIOD: 001:00:00 NO. OF RECOR

Page 38

Operating Instructions Teledyne API – T101 Operation Manual 130 4.8.2.1. Viewing DAS Data and Settings DAS data and settings can be viewed on the

Page 39

Teledyne API – T101 Operation Manual Operating Instructions 131 4.8.2.2. Editing DAS Data Channels DAS configuration is most conveniently done throu

Page 40

Operating Instructions Teledyne API – T101 Operation Manual 132 PARAMETERS: Four parameters are included in this channel EVENT: This channel is se

Page 41 - 3.4. PNEUMATIC CONNECTIONS

Teledyne API – T101 Operation Manual Operating Instructions 133 4.8.2.3. Trigger Events To edit the list of data parameters associated with a specif

Page 42

Operating Instructions Teledyne API – T101 Operation Manual 134 4.8.2.4. Editing DAS Parameters Data channels can be edited individually from the

Page 43

Teledyne API – T101 Operation Manual Operating Instructions 135 To configure the parameters for a specific data parameter, press: FROM THE EDIT DA

Page 44 - CO NTR O L

Operating Instructions Teledyne API – T101 Operation Manual 136 SAMPLE PERIOD is only used when the DAS parameter’s sample mode is set for AVG, MI

Page 45

Teledyne API – T101 Operation Manual Operating Instructions 137 To define the REPORT PERIOD, follow the instruction shown in Section 4.11.2.2 then p

Page 46

TABLE OF CONTENTS Teledyne API – T101 Operation Manual xii 7.5. Dynamic Multipoint Span Calibration ...

Page 47

Operating Instructions Teledyne API – T101 Operation Manual 138 combined number of records would be exceeded. (For information regarding the maxim

Page 48

Teledyne API – T101 Operation Manual Operating Instructions 139 4.8.2.7. RS-232 Report Function The T101 DAS can automatically report data to the co

Page 49

Operating Instructions Teledyne API – T101 Operation Manual 140 4.8.2.10. Disabling/Enabling Data Channels Data channels can be temporarily disabl

Page 50 - Permeation Tube Heater

Teledyne API – T101 Operation Manual Operating Instructions 141 4.8.2.11. HOLDOFF Feature The DAS HOLDOFF feature allows to prevent data collection

Page 51 - NO

Operating Instructions Teledyne API – T101 Operation Manual 142 4.8.3. REMOTE DAS CONFIGURATION Editing channels, parameters and triggering events

Page 52 - CALIBRATION

Teledyne API – T101 Operation Manual Remote Operation 143 5. REMOTE OPERATION 5.1.1. REMOTE OPERATION USING THE EXTERNAL DIGITAL I/O 5.1.1.1. Stat

Page 53

Remote Operation Teledyne API – T101 Operation Manual 144 Connect to Internal Ground of MonitoringSTATUS 1 2 3 4 5 6

Page 54 - 3.5.4. FUNCTIONAL CHECK

Teledyne API – T101 Operation Manual Remote Operation 145 Table 5-1. Status Output Pin Assignments CONNECTOR PIN STATUS CONDITION (ON=CONDUCTING)

Page 55 - 3.6. INITIAL CALIBRATION

Remote Operation Teledyne API – T101 Operation Manual 146 There are two methods to activate control inputs. The internal +5V available from the “+

Page 56

Teledyne API – T101 Operation Manual Remote Operation 147 5.1.2.2. Help Commands in Terminal Mode Table 5-3. Terminal Mode Software Commands COMMAN

Page 57

Teledyne API – T101 Operation Manual TABLE OF CONTENTS xiii 9.5.12. PMT Sensor ...

Page 58 - S MEASUREMENTS

Remote Operation Teledyne API – T101 Operation Manual 148 5.1.2.3. Command Syntax Commands are not case-sensitive and all arguments within one com

Page 59 - 4. OPERATING INSTRUCTIONS

Teledyne API – T101 Operation Manual Remote Operation 149 hexadecimal digits (0-9, A-F, a-f), which is the ‘C’ programming language convention. No

Page 60 - 4.2. SAMPLE MODE

Remote Operation Teledyne API – T101 Operation Manual 150 5.1.2.6. General Message Format All messages from the instrument (including those in res

Page 61

Teledyne API – T101 Operation Manual Remote Operation 151 To change this setting press: SETUP X.X SECONDARY SETUP MENU COMM VARS DIAG

Page 62

Remote Operation Teledyne API – T101 Operation Manual 152 To Initialize the modem press: Select which COM Port is tested SETUP X.X SECONDA

Page 63 - 4.2.2. WARNING MESSAGES

Teledyne API – T101 Operation Manual Remote Operation 153 5.1.2.8. COM Port Password Security In order to provide security for remote access of the

Page 64 - 4.3. CALIBRATION MODE

Remote Operation Teledyne API – T101 Operation Manual 154 Figure 5-4. APICOM Remote Control Program Interface APICOM is included free of cost wit

Page 65

Teledyne API – T101 Operation Manual Remote Operation 155 5.1.4. USING THE T101 WITH A HESSEN PROTOCOL NETWORK 5.1.4.1. General Overview of Hessen

Page 66 - 4.4. SETUP MODE

Remote Operation Teledyne API – T101 Operation Manual 156 5.1.4.3. Activating Hessen Protocol The first step in configuring the T101 to operate ov

Page 67

Teledyne API – T101 Operation Manual Remote Operation 157 5.1.4.4. Selecting a Hessen Protocol Type Currently there are two version of Hessen Proto

Page 68

TABLE OF CONTENTS Teledyne API – T101 Operation Manual xiv 10.4.10.1. A to D Conversion ...

Page 69

Remote Operation Teledyne API – T101 Operation Manual 158 To Select a Hessen response mode, press: SETUP X.X SECONDARY SETUP MENU COMM VA

Page 70

Teledyne API – T101 Operation Manual Remote Operation 159 5.1.4.6. Hessen Protocol Gas ID Since the T101 can be, when the proper optional equipment

Page 71

Remote Operation Teledyne API – T101 Operation Manual 160 5.1.4.7. Setting Hessen Protocol Status Flags Teledyne API’s implementation of Hessen pr

Page 72

Teledyne API – T101 Operation Manual Remote Operation 161 To assign or reset the status flag bit assignments, press: SETUP X.X SECONDARY SETUP

Page 73

Remote Operation Teledyne API – T101 Operation Manual 162 This page intentionally left blank. 07266B DCN6485

Page 74 - 4.4.4.7. Range Units

Teledyne API – T101 Operation Manual Calibration Procedures 163 6. CALIBRATION PROCEDURES This Section describes the calibration procedures for the

Page 75 - 4.4.4.8. Dilution Ratio

Calibration Procedures Teledyne API – T101 Operation Manual 164 6.1.2. ZERO AIR Zero air is similar in chemical composition to the Earth’s atmosph

Page 76 - Table 4-6. Password Levels

Teledyne API – T101 Operation Manual Calibration Procedures 165 6.1.5. CALIBRATION GAS TRACEABILITY All equipment used to produce calibration gases

Page 77 - PRIMARY SETUP MENU

Calibration Procedures Teledyne API – T101 Operation Manual 166 STEP ONE: Connect the sources of zero air and span gas as shown below. No Valve O

Page 78 - SECONDARY SETUP MENU

Teledyne API – T101 Operation Manual Calibration Procedures 167 STEP TWO: Set the expected H2S span gas concentrations. In this example the inst

Page 79

Teledyne API – T101 Operation Manual TABLE OF CONTENTS xv LIST OF FIGURES Figure 3-1. Front Panel Layout ...

Page 80 - ALLOWED

Calibration Procedures Teledyne API – T101 Operation Manual 168 STEP THREE: Perform the zero/span calibration: M-P CAL H2S STB=X.XX

Page 81 - See Section 6.8.1

Teledyne API – T101 Operation Manual Calibration Procedures 169 6.3. MANUAL CALIBRATION CHECKS Informal calibration checks, which only evaluate but

Page 82 - Press the PREV

Calibration Procedures Teledyne API – T101 Operation Manual 170 6.4. MANUAL CALIBRATION WITH ZERO/SPAN VALVES Zero and Span calibrations using the

Page 83 - FUNCTIONS

Teledyne API – T101 Operation Manual Calibration Procedures 171 Step Two: Set the expected H2S span gas value: This seque nce ca uses the analyze

Page 84

Calibration Procedures Teledyne API – T101 Operation Manual 172 Step Three: Perform the calibration or calibration check according to the followin

Page 85 - 4.6.1. SIGNAL I/O

Teledyne API – T101 Operation Manual Calibration Procedures 173 6.5. MANUAL CALIBRATION WITH IZS OPTION The following section describes the basic m

Page 86

Calibration Procedures Teledyne API – T101 Operation Manual 174 6.6. MANUAL CALIBRATION CHECKS WITH IZS OR ZERO/SPAN VALVES Zero and span checks u

Page 87

Teledyne API – T101 Operation Manual Calibration Procedures 175 STEP ONE: Connect the sources of Zero Air and Span Gas as shown below. Exter

Page 88 - Table 4-12

Calibration Procedures Teledyne API – T101 Operation Manual 176 STEP TWO: Perform the zero/span check. Wait until H2S STB falls below 0.5 pp b.

Page 89

Teledyne API – T101 Operation Manual Calibration Procedures 177 6.7. MANUAL CALIBRATION IN INDEPENDENT OR AUTO REPORTING RANGE MODES If the analyze

Page 90

TABLE OF CONTENTS Teledyne API – T101 Operation Manual xvi Figure 10-8. Typical Flow Control Assembly with Critical Flow Orifice ...

Page 91

Calibration Procedures Teledyne API – T101 Operation Manual 178 If contact closures are used in conjunction with the analyzer’s AutoCal (Section 6

Page 92

Teledyne API – T101 Operation Manual Calibration Procedures 179 The analyzer will ask to have the GAS TYPE specified at the beginning of the proces

Page 93 - 4.6.3), then press:

Calibration Procedures Teledyne API – T101 Operation Manual 180 Table 6-3. AutoCal Attribute Setup Parameters Attribute Name ACTION Timer Enabled

Page 94

Teledyne API – T101 Operation Manual Calibration Procedures 181 The following example sets Sequence2 to carry out a zero-span calibration every oth

Page 95 - (Figure 4-5)

Calibration Procedures Teledyne API – T101 Operation Manual 182 Toggle to se t number of days between procedures (1-367) Toggle to set delay time

Page 96

Teledyne API – T101 Operation Manual Calibration Procedures 183 It is recommended that, if dynamic calibration is enabled, the analyzer’s test func

Page 97 - 4.6.3.6. AIN Calibration

Calibration Procedures Teledyne API – T101 Operation Manual 184 6.9.2. AUTOCAL OF INSTRUMENTS IN MULTIGAS MEASUREMENT MODE If the analyzer is bein

Page 98

Teledyne API – T101 Operation Manual Calibration Procedures 185 6.10. CALIBRATION QUALITY After completing one of the calibration procedures descri

Page 99 - 4.6.4. OPTIC TEST

Calibration Procedures Teledyne API – T101 Operation Manual 186 This page intentionally left blank. 07266B DCN6485

Page 100 - 4.6.5. ELECTRICAL TEST

187 7. EPA PROTOCOL CALIBRATION 7.1. CALIBRATION REQUIREMENTS When configured for SO2 measurement mode or multigas measurement mode, the T101 may b

Page 101 - 4.6.6. LAMP CALIBRATION

Teledyne API – T101 Operation Manual TABLE OF CONTENTS xvii Table 4-14. Current Loop Output Calibration with Resistor ...

Page 102 - 4.6.7. PRESSURE CALIBRATION

EPA Protocol Calibration Teledyne API – T101 Operation Manual 188 time. Therefore, it is necessary to dynamically check the calibration relationsh

Page 103 - 4.6.8. FLOW CALIBRATION

Teledyne API – T101 Operation Manual EPA Protocol Calibration 189 Table 7-2. Activity Matrix for Calibration Procedure EQUIPMENT/SUPPLIES ACCEPTANC

Page 104 - 4.6.9. TEST CHANNEL OUTPUT

EPA Protocol Calibration Teledyne API – T101 Operation Manual 190 7.1.5. CALIBRATION FREQUENCY To ensure accurate measurements of the SO2 concentr

Page 105 - COMMUNICATION PORTS

Teledyne API – T101 Operation Manual EPA Protocol Calibration 191 7.1.7. SUMMARY OF QUALITY ASSURANCE CHECKS The following items should be checked

Page 106

EPA Protocol Calibration Teledyne API – T101 Operation Manual 192 Table 7-4. Definition of Level 1 and Level 2 Zero and Span Checks (from Section

Page 107

Teledyne API – T101 Operation Manual EPA Protocol Calibration 193 7.3. ZERO AND SPAN CHECKS A system of Level 1 and Level 2 zero span checks (see

Page 108

EPA Protocol Calibration Teledyne API – T101 Operation Manual 194  Single range selection. See Section 4.7.4 of this manual. If the instrument w

Page 109 - 4.7.4. RS-485 CONFIGURATION

Teledyne API – T101 Operation Manual EPA Protocol Calibration 195 7.5. DYNAMIC MULTIPOINT SPAN CALIBRATION Dynamic calibration involves introducing

Page 110

EPA Protocol Calibration Teledyne API – T101 Operation Manual 196 7.6. SPECIAL CALIBRATION REQUIREMENTS FOR INDEPENDENT RANGE OR AUTO RANGE If Ind

Page 111

197 8. INSTRUMENT MAINTENANCE Predictive diagnostic functions including data acquisition, failure warnings and alarms built into the analyzer allow

Page 113 - Network Administrator

TABLE OF CONTENTS Teledyne API – T101 Operation Manual xviii This page intentionally left blank. 07266B DCN6485

Page 114

Instrument Maintenance Teledyne API – T101 Operation Manual 198 047400102 Rev XC This page intentionally left blank. 07266B DCN6485

Page 115 - >\

Instrument Maintenance Model T101 Instruction Manual 199 Table 8-1 T101 Preventive Maintenance Schedule ITEM ACTION FREQUENCY CAL CHECK MANUAL SE

Page 116 - 4.7.7. USB CONFIGURATION

Instrument Maintenance Model T101 Instruction Manual 200 This page intentionally left blank. 07266B DCN6485

Page 117

Model T101 Instruction Manual Instrument Maintenance 201 8.2. PREDICTIVE DIAGNOSTICS The analyzer’s test functions can be used to predict failure

Page 118

Instrument Maintenance Model T101 Instruction Manual 202 8.3. MAINTENANCE PROCEDURES The following procedures need to be performed regularly as part

Page 119

Model T101 Instruction Manual Instrument Maintenance 203 8.3.2. CHANGING THE IZS PERMEATION TUBE 1. Turn off the analyzer, unplug the power cord a

Page 120 - RS-232 port

Instrument Maintenance Model T101 Instruction Manual 204 EXAMPLE: If the SO2 concentration is 125 ppb: Operational hours = 1000 ppm/hr ÷ 0.125 pp

Page 121 - 4.7.9. MODBUS SET UP

Model T101 Instruction Manual Instrument Maintenance 205 8.3.4. CHANGING THE EXTERNAL ZERO AIR SCRUBBER The chemicals in the external scrubber need

Page 122

Instrument Maintenance Model T101 Instruction Manual 206 8.3.5. MAINTAINING THE H2S  SO2 CONVERTER The catalyst contained in the H2S  SO2 converte

Page 123 - ID DESCRIPTION

Model T101 Instruction Manual Instrument Maintenance 207 Converter Efficiency = 490.3 ÷ 499.1 Converter Efficiency = 0.982 (98.2%) 10. It is

Page 124

19 1. INTRODUCTION The Model T101 UV Fluorescence H2S Analyzer measures hydrogen sulfide in levels commonly required for Ambient Air monitoring. Th

Page 125 - 4.7.11. COM PORT BAUD RATE

Instrument Maintenance Model T101 Instruction Manual 208 Converter Assembly Cover Band Heater and T/C Assembly H2S  SO2 converter Converter Assembly

Page 126 - 4.7.12. COM PORT TESTING

Model T101 Instruction Manual Instrument Maintenance 209 8.3.6. CHECKING FOR LIGHT LEAKS When re-assembled or operated improperly, the T101 can dev

Page 127 - 4.8.1. DAS STRUCTURE

Instrument Maintenance Model T101 Instruction Manual 210 Gas Line fitting Spring Sintered Filter O-Ring Critical Flow Orifice O-Ring Vacuum Manifold

Page 128 - 4.8.1.2. DAS Parameters

Model T101 Instruction Manual Troubleshooting & Service 211 9. TROUBLESHOOTING & SERVICE This section contains a variety of methods for i

Page 129

Troubleshooting & Service Model T101 Instruction Manual 212  Examine the values of all TEST functions and compare them to factory values. Note

Page 130 - 4.8.2. DEFAULT DAS CHANNELS

Model T101 Instruction Manual Troubleshooting & Service 213 (Fault LED lit, warning msg in Param field, CLR button active) (Fault LED lit, Te

Page 131 - LIST OF CHANNELS

Troubleshooting & Service Model T101 Instruction Manual 214 To view or clear a warning message press: SAMPLE SYSTEM RESET H

Page 132

Model T101 Instruction Manual Troubleshooting & Service 215 WARNING MESSAGE FAULT CONDITION POSSIBLE CAUSES IZS TEMP WARNING On units with IZS

Page 133 - 0) CONC: ATIMER, 4, 800

Troubleshooting & Service Model T101 Instruction Manual 216 9.1.2. FAULT DIAGNOSIS WITH TEST FUNCTIONS Besides being useful as predictive diagnos

Page 134

Model T101 Instruction Manual Troubleshooting & Service 217 9.1.3. USING THE DIAGNOSTIC SIGNAL I/O FUNCTION The signal I/O parameters found und

Page 135 - 4.8.2.3. Trigger Events

Introduction Teledyne API – T101 Operation Manual 20 1.2. OPTIONS The options available for your analyzer are presented in Table 1-1 with name, op

Page 136 - Press SET> until…

Troubleshooting & Service Model T101 Instruction Manual 218 Toggles parameter ON/OFFExit returns to DIAG display & all values return to

Page 137

Model T101 Instruction Manual Troubleshooting & Service 219 9.1.4.1. Motherboard Status Indicator (Watchdog) DS5, a red LED on the upper portio

Page 138

Troubleshooting & Service Model T101 Instruction Manual 220 Table 9-3. Relay Board Status LEDs LED COLOR FUNCTION FAULT STATUS INDICATED FAILUR

Page 139 - 4.8.2.6. Number of Records

Model T101 Instruction Manual Troubleshooting & Service 221 9.2.1. ZERO OR LOW SAMPLE FLOW If the pump is operating but the unit reports a 0 ga

Page 140

Troubleshooting & Service Model T101 Instruction Manual 222  If the response offset test function for H2S (H2S OFFS) are greater than 150 mV, a

Page 141 - 4.8.2.9. Starting Date

Model T101 Instruction Manual Troubleshooting & Service 223  Verify that the expected concentration is set properly to the actual span gas co

Page 142

Troubleshooting & Service Model T101 Instruction Manual 224  Dilution air contains sample or span gas.  Sample inlet may be contaminated with

Page 143 - 4.8.2.11. HOLDOFF Feature

Model T101 Instruction Manual Troubleshooting & Service 225  Dirty sample chamber. Clean the sample chamber.  Insufficient time allowed for

Page 144

Troubleshooting & Service Model T101 Instruction Manual 226 bubble solution, looking for fine bubbles. Once the fittings have been wetted with so

Page 145 - 5. REMOTE OPERATION

Model T101 Instruction Manual Troubleshooting & Service 227 If the unit is set for 220-240 V and is plugged into 100-120 V, the analyzer will n

Page 146

Teledyne API – T101 Operation Manual Introduction 21 Option Option Number Description/Notes Analog Inputs Used for connecting external voltage sign

Page 147 - 5.1.1.2. Control Inputs

Troubleshooting & Service Model T101 Instruction Manual 228 9.5.5. I2C BUS Operation of the I2C bus can be verified by observing the behavior of

Page 148

Model T101 Instruction Manual Troubleshooting & Service 229 Table 9-6. Relay Board Control Devices FUNCTION CONTROL DEVICE SOCKETED Valve0 – Va

Page 149

Troubleshooting & Service Model T101 Instruction Manual 230 9.5.9.3. Status Outputs The procedure below can be used to test the Status outputs.

Page 150 - 5.1.2.4. Data Types

Model T101 Instruction Manual Troubleshooting & Service 231 In some rare circumstances, this failure may be caused by a bad IC on the motherboa

Page 151 - 5.1.2.5. Status Reporting

Troubleshooting & Service Model T101 Instruction Manual 232 Further help with serial communications is available in a separate manual “RS-232 Man

Page 152

Model T101 Instruction Manual Troubleshooting & Service 233 9.5.15. HIGH VOLTAGE POWER SUPPLY The HVPS is located in the interior of the sensor

Page 153

Troubleshooting & Service Model T101 Instruction Manual 234 The IZS option is heated with a proportional heater circuit and the temperature is ma

Page 154

Model T101 Instruction Manual Troubleshooting & Service 235 be re-entered before the instrument will function correctly. Also, zero and span ca

Page 155

Troubleshooting & Service Model T101 Instruction Manual 236 Figure 9-4. Shutter Assembly CAUTION ALWAYS wear UV-Protective, Safety Glasses when

Page 156

Model T101 Instruction Manual Troubleshooting & Service 237 Table 9-9. Example of UV Lamp Power Supply Outputs UVLAMP_SIGNAL ACTION TO BE TAKE

Page 157 - Duplex Full Half

Introduction Teledyne API – T101 Operation Manual 22 Option Option Number Description/Notes RS-232 Multidrop Enables communications between host c

Page 158

Troubleshooting & Service Model T101 Instruction Manual 238 NOTE DO NOT grasp the UV lamp by its cap when changing its position (refer to Figure

Page 159

Model T101 Instruction Manual Troubleshooting & Service 239 Figure 9-6. Pre-Amplifier Board Layout 6. Turn the gain adjustment potentiometer

Page 160

Troubleshooting & Service Model T101 Instruction Manual 240 9.7. FREQUENTLY ASKED QUESTIONS (FAQS) The following list contains some of the most c

Page 161

Model T101 Instruction Manual Troubleshooting & Service 241 QUESTION ANSWER How often do I need to change the particulate filter? Once per week

Page 162

Troubleshooting & Service Model T101 Instruction Manual 242 This page intentionally left blank. 07266B DCN6485

Page 163

Model T101 Instruction Manual Principles Of Operation 243 10. PRINCIPLES OF OPERATION The T101 UV Fluorescence H2S Analyzer is a microprocessor c

Page 164

Principles Of Operation Model T101 Instruction Manual 244 The chemical process is: 2222SO2OH2O3SH2  The converter is a heated stainless steel c

Page 165 - 6. CALIBRATION PROCEDURES

Model T101 Instruction Manual Principles Of Operation 245 Figure 10-1. UV Absorption The equation for defining the average intensity of the UV lig

Page 166 - 6.1.4. PERMEATION TUBES

Principles Of Operation Model T101 Instruction Manual 246 *SOkF2 Where: F = the amount of fluorescent light given off. k = The rate at whic

Page 167 - 6.2. MANUAL CALIBRATION

Model T101 Instruction Manual Principles Of Operation 247 10.2. THE UV LIGHT PATH The optical design of the Model T101’s sample chamber optimizes t

Page 168 - Chassis

23 2. SPECIFICATIONS AND APPROVALS 2.1. SPECIFICATIONS Table 2-1. Model T101 Basic Unit Specifications PARAMETER DESCRIPTION Ranges H2S: Min 0-50 p

Page 169 - STEP TWO: Set the expected H

Principles Of Operation Model T101 Instruction Manual 248 The lamp used in the Model T101 is constructed with a vacuum jacket surrounding a double-bo

Page 170

Model T101 Instruction Manual Principles Of Operation 249 10.2.4. OPTICAL FILTERS The Model T101 analyzer uses two stages of optical filters to enh

Page 171

Principles Of Operation Model T101 Instruction Manual 250 LAMP OUTPUT (Arbitrary Untis) 1051041021031011100 200 300 400 500 WAVELENGTH (nm) 0213.9 3

Page 172 - (At high

Model T101 Instruction Manual Principles Of Operation 251 Another lens located between the excitation UV source lamp and the sample chamber collima

Page 173 - span gas value:

Principles Of Operation Model T101 Instruction Manual 252 10.2.6.2. UV Absorption by Ozone Because ozone absorbs UV Light over a relatively broad spe

Page 174

Model T101 Instruction Manual Principles Of Operation 253 10.3. PNEUMATIC OPERATION CAUTION It is important that the sample airflow system is leak-

Page 175

Principles Of Operation Model T101 Instruction Manual 254 10.3.1. SAMPLE GAS FLOW The flow of gas through the T101 UV Fluorescence H2S Analyzer is cr

Page 176 - ZERO/SPAN VALVES

Model T101 Instruction Manual Principles Of Operation 255 10.3.2. MULTIGAS MEASUREMENT & H2S  SO2 SWITCHING VALVE. When activated for operatio

Page 177

Principles Of Operation Model T101 Instruction Manual 256 pressure differential combined with the action of the analyzer’s external pump draws the ga

Page 178

Model T101 Instruction Manual Principles Of Operation 257 10.3.5. HYDROCARBON SCRUBBER (KICKER) It is very important to make sure the air supplied

Page 179

Specifications and Approvals Teledyne API – T101 Operation Manual 24 PARAMETER DESCRIPTION Standard I/O 1 Ethernet: 10/100Base-T 2 RS-232 (300 –

Page 180 - No Valve Options Installed

Principles Of Operation Model T101 Instruction Manual 258 10.3.7. PNEUMATIC SENSORS The T101 uses two pneumatic sensors to verify gas streams. These

Page 181 - Table 6-2. AutoCal Modes

Model T101 Instruction Manual Principles Of Operation 259 10.4. ELECTRONIC OPERATION Analog In Pneumatic Sensor Board Sample Pressure Sensor Samp

Page 182

Principles Of Operation Model T101 Instruction Manual 260 The core of the analyzer is a microcomputer that controls various internal processes, inter

Page 183 - 0.5 hours later each day

Model T101 Instruction Manual Principles Of Operation 261 10.4.1. CPU The unit’s CPU card, installed on the motherboard located inside the rear pan

Page 184 - Default

Principles Of Operation Model T101 Instruction Manual 262 10.4.2. SENSOR MODULE & SAMPLE CHAMBER Electronically, the T101 sensor module is a grou

Page 185 - REPORTING RANGE MODES

Model T101 Instruction Manual Principles Of Operation 263 10.4.4. PHOTO MULTIPLIER TUBE (PMT) The T101 uses a photo multiplier tube (PMT) to detect

Page 186 - MEASUREMENT MODE

Principles Of Operation Model T101 Instruction Manual 264 Figure 10-14. Basic PMT Design A significant performance characteristic of the PMT is the

Page 187 - 6.10. CALIBRATION QUALITY

Model T101 Instruction Manual Principles Of Operation 265 TEC Control PCA PMT Preamp PCA Thermistor outputs temp of cold block to preamp PCA Pr

Page 188

Principles Of Operation Model T101 Instruction Manual 266 be adjusted through a potentiometer. These adjustments should only be performed when encoun

Page 189 - 7. EPA PROTOCOL CALIBRATION

Model T101 Instruction Manual Principles Of Operation 267 10.4.7. PNEUMATIC SENSOR BOARD The flow and pressure sensors of the T101 are located on a

Page 190

25 3. GETTING STARTED 3.1. UNPACKING AND INITIAL SETUP CAUTION To avoid personal injury, always use two persons to lift and carry the Model T101.

Page 191 - TRACEABILITY

Principles Of Operation Model T101 Instruction Manual 268 10.4.9. STATUS LEDS & WATCH DOG CIRCUITRY Zero/Span and IZS Options Sample/Cal Valve I

Page 192 - 7.1.6. RECORD KEEPING

Model T101 Instruction Manual Principles Of Operation 269 As a Safety measure, special circuitry on the Relay Board watches the status of LED D1. S

Page 193 - Table 7-3. Activity Matrix

Principles Of Operation Model T101 Instruction Manual 270 board. It is digitized and sent to the CPU where it is used to calculate the current temper

Page 194

Model T101 Instruction Manual Principles Of Operation 271 In its standard configuration, the analyzer comes with all three of these channels set up

Page 195 - 7.3. ZERO AND SPAN CHECKS

Principles Of Operation Model T101 Instruction Manual 272 AC POWER ENTRANCE ON/OFF SWITCH PS 1 (+5 VDC; ±15 VDC) PS 2 (+12 VDC) Chassis Cooling Fa

Page 196 - 7.4.2. PRECISION CHECK

Model T101 Instruction Manual Principles Of Operation 273 10.6. FRONT PANEL/DISPLAY INTERFACE Users can input data and receive information directly

Page 197

Principles Of Operation Model T101 Instruction Manual 274 10.7. SOFTWARE OPERATION The T101 H2S analyzer is at its heart a high performance, 386-base

Page 198 - 7.7. REFERENCES

Model T101 Instruction Manual Principles Of Operation 275 10.7.2. CALIBRATION - SLOPE AND OFFSET Calibration of the analyzer is performed exclusive

Page 199 - 8. INSTRUMENT MAINTENANCE

Principles Of Operation Model T101 Instruction Manual 276 10.7.3. TEMPERATURE AND PRESSURE COMPENSATION (TPC) FEATURE As explained previously, change

Page 200 - 198 047400102 Rev XC

Model T101 Instruction Manual A Primer on Electro-Static Discharge 277 11. A PRIMER ON ELECTRO-STATIC DISCHARGE Teledyne API considers the preven

Page 201

Getting Started Teledyne API – T101 Operation Manual 26 4. Inspect the interior of the instrument to make sure all circuit boards and other compo

Page 202

A Primer on Electro-Static Discharge Model T101 Instruction Manual 278 workbench, using a plastic handled screwdriver or even the constant jostling o

Page 203

Model T101 Instruction Manual A Primer on Electro-Static Discharge 279  Any time a charged surface (including the human body) discharges to a dev

Page 204 - 8.3. MAINTENANCE PROCEDURES

A Primer on Electro-Static Discharge Model T101 Instruction Manual 280  As long as my analyzer is properly installed, it is safe from damage caused

Page 205 - SCRUBBER

Model T101 Instruction Manual A Primer on Electro-Static Discharge 281  Always store sensitive components and assemblies in anti-ESD storage bags

Page 206 - To check to see if your SO

A Primer on Electro-Static Discharge Model T101 Instruction Manual 282 11.4.2. BASIC ANTI-ESD PROCEDURES FOR ANALYZER REPAIR AND MAINTENANCE 11.4.2.1

Page 207

Model T101 Instruction Manual A Primer on Electro-Static Discharge 283 6. Disconnecting your wrist strap is always the last action taken before le

Page 208 - CONVERTER

A Primer on Electro-Static Discharge Model T101 Instruction Manual 284 static charges from damaging the components or assemblies being shipped make s

Page 209 - 8.3.5.3. Changing the H

Model T101 Instruction Manual Clossary 285 GLOSSARY Note: Some terms in this glossary may not occur elsewhere in this manual. Term Description/Def

Page 210

Glossary Model T101 Instruction Manual 286 Term Description/Definition O2 molecular oxygen O3 ozone SO2 sulfur dioxide cm3 metric abbreviation for c

Page 211

Model T101 Instruction Manual Glossary 287 Term Description/Definition FPI Fabry-Perot Interface: a special light filter typically made of a trans

Page 212

Teledyne API – T101 Operation Manual Getting Started 27 Figure 3-1. Front Panel Layout Figure 3-2. Display Screen and Touch Control CAUTION –

Page 213 - 9.1. GENERAL TROUBLESHOOTING

Glossary Model T101 Instruction Manual 288 Term Description/Definition PFA Per-Fluoro-Alkoxy, an inert polymer; one of the polymers that Du Pont mar

Page 214

Teledyne API T/E-Series Models 101, 102, and 108 Software Documentation (PN05492D DCN6485) A-1 APPENDIX A - Version Specific Software Documentation

Page 215

Teledyne API T/E-Series Models 101, 102, and 108 Software Documentation (PN05492D DCN6485) A-2 This page intentionally left blank. 07266B DCN6485

Page 216

Teledyne API T/E-Series Models 101, 102, and 108 Software Documentation (PN05492D DCN6485) A-3 APPENDIX A-1: Software Menu Trees, S/W Version C.7

Page 217

Teledyne API T/E-Series Models 101, 102, and 108 Software Documentation (PN05492D DCN6485) A-4 LOW HIGH ZERO SPAN CONC SAMPLE <TST TST> SET

Page 218

Teledyne API T/E-Series Models 101, 102, and 108 Software Documentation (PN05492D DCN6485) A-5 CFG  MODEL NAME  PART NUMBER  SERIAL NUMBER

Page 219

Teledyne API T/E-Series Models 101, 102, and 108 Software Documentation (PN05492D DCN6485) A-6 NAME EVENT PARAMETERS REPORT PERIOD NUMBER OF RECOR

Page 220 - 9.1.4. STATUS LEDS

Teledyne API T/E-Series Models 101, 102, and 108 Software Documentation (PN05492D DCN6485) A-7 QUIET COMPUTER HESSEN PROTOCOL E, 8,1 E, 7, 1 RS

Page 221 - CPU Status LED

Teledyne API T/E-Series Models 101, 102, and 108 Software Documentation (PN05492D DCN6485) A-8 INSTRUMENT IP5 GATEWAY IP5 SUBNET MASK5 TCP PORT3 D

Page 222 - 9.2. GAS FLOW PROBLEMS

Teledyne API T/E-Series Models 101, 102, and 108 Software Documentation (PN05492D DCN6485) A-9 . YES NO SAMPLE FLOW WARNING BENCH TEMP WARNI

Page 223 - 9.3. CALIBRATION PROBLEMS

i ABOUT TELEDYNE ADVANCED POLLUTION INSTRUMENTATION (TAPI) Teledyne Advanced Pollution Instrumentation (TAPI), a business unit of Teledyne Instrume

Page 224 - 9.3.2. NO RESPONSE

Getting Started Teledyne API – T101 Operation Manual 28 The front panel liquid crystal display screen includes touch control. Upon analyzer start-

Page 225 - 9.3.6. NON-LINEAR RESPONSE

Teledyne API T/E-Series Models 101, 102, and 108 Software Documentation (PN05492D DCN6485) A-10 37 INTERNAL ANALOG  VOLTAGE SIGNALS 72 (see Ap

Page 226 - 9.4.2. SLOW RESPONSE

Teledyne API T/E-Series Models 101, 102, and 108 Software Documentation (PN05492D DCN6485) A-11 APPENDIX A-2: Setup Variables For Serial I/O, S/W Vers

Page 227 - 9.5. SUBSYSTEM CHECKOUT

Teledyne API T/E-Series Models 101, 102, and 108 Software Documentation (PN05492D DCN6485) A-12 Setup Variable Numeric Units Default Value Value Ran

Page 228

Teledyne API T/E-Series Models 101, 102, and 108 Software Documentation (PN05492D DCN6485) A-13 Setup Variable Numeric Units Default Value Value Ran

Page 229 - 9.5.4. DC POWER SUPPLY

Teledyne API T/E-Series Models 101, 102, and 108 Software Documentation (PN05492D DCN6485) A-14 Setup Variable Numeric Units Default Value Value Ran

Page 230 - 9.5.7. LCD DISPLAY MODULE

Teledyne API T/E-Series Models 101, 102, and 108 Software Documentation (PN05492D DCN6485) A-15 Setup Variable Numeric Units Default Value Value Ran

Page 231 - 9.5.9. MOTHERBOARD

Teledyne API T/E-Series Models 101, 102, and 108 Software Documentation (PN05492D DCN6485) A-16 Setup Variable Numeric Units Default Value Value Ran

Page 232 - 9.5.10. CPU

Teledyne API T/E-Series Models 101, 102, and 108 Software Documentation (PN05492D DCN6485) A-17 Setup Variable Numeric Units Default Value Value Ran

Page 233 - 9.5.11. RS-232 COMMUNICATION

Teledyne API T/E-Series Models 101, 102, and 108 Software Documentation (PN05492D DCN6485) A-18 Setup Variable Numeric Units Default Value Value Ran

Page 234 - 9.5.12. PMT SENSOR

Teledyne API T/E-Series Models 101, 102, and 108 Software Documentation (PN05492D DCN6485) A-19 Setup Variable Numeric Units Default Value Value Ran

Page 235 - 9.5.17. IZS OPTION

Teledyne API – T101 Operation Manual Getting Started 29 Figure 3-3. Display/Touch Control Screen Mapped to Menu Charts 07266B DCN6485

Page 236 - 9.6. REPAIR PROCEDURES

Teledyne API T/E-Series Models 101, 102, and 108 Software Documentation (PN05492D DCN6485) A-20 Setup Variable Numeric Units Default Value Value Ran

Page 237

Teledyne API T/E-Series Models 101, 102, and 108 Software Documentation (PN05492D DCN6485) A-21 APPENDIX A-3: Warnings and Test Measurements, S/W Vers

Page 238

Teledyne API T/E-Series Models 101, 102, and 108 Software Documentation (PN05492D DCN6485) A-22 WPMTTEMP PMT TEMP WARNING PMT temperature outside

Page 239 - 9.6.3. REPLACING THE UV LAMP

Teledyne API T/E-Series Models 101, 102, and 108 Software Documentation (PN05492D DCN6485) A-23 Table A-3: Test Measurements TEST Measurement M

Page 240

Teledyne API T/E-Series Models 101, 102, and 108 Software Documentation (PN05492D DCN6485) A-24 TEST Measurement Message Text DESCRIPTION PMTTEMP

Page 241

Teledyne API T/E-Series Models 101, 102, and 108 Software Documentation (PN05492D DCN6485) A-25 APPENDIX A-4: Signal I/O Definitions, S/W Version C.7

Page 242

Teledyne API T/E-Series Models 101, 102, and 108 Software Documentation (PN05492D DCN6485) A-26 Signal Name Bit or Channel Number Description 6–7 A

Page 243 - 9.8. TECHNICAL ASSISTANCE

Teledyne API T/E-Series Models 101, 102, and 108 Software Documentation (PN05492D DCN6485) A-27 Signal Name Bit or Channel Number Description 1 = al

Page 244

Teledyne API T/E-Series Models 101, 102, and 108 Software Documentation (PN05492D DCN6485) A-28 Signal Name Bit or Channel Number Description DARK_S

Page 245 - 10. PRINCIPLES OF OPERATION

Teledyne API T/E-Series Models 101, 102, and 108 Software Documentation (PN05492D DCN6485) A-29 Signal Name Bit or Channel Number Description TEMP_I

Page 246 - SO2OH2O3SH2 

Getting Started Teledyne API – T101 Operation Manual 30 3.2.2. REAR PANEL Figure 3-4. Rear Panel Layout 07266B DCN6485

Page 247 - SOaxexp1IIa

Teledyne API T/E-Series Models 101, 102, and 108 Software Documentation (PN05492D DCN6485) A-30 Table A-5: DAS Trigger Events Name Description

Page 248 - = Amount of excited state SO

Teledyne API T/E-Series Models 101, 102, and 108 Software Documentation (PN05492D DCN6485) A-31 Appendix A-6: DAS Functions Name Description Units

Page 249 - 10.2.1. UV SOURCE LAMP

Teledyne API T/E-Series Models 101, 102, and 108 Software Documentation (PN05492D DCN6485) A-32 Name Description Units RCTEMP Reaction cell temper

Page 250 - 10.2.3. THE PMT

Teledyne API T/E-Series Models 101, 102, and 108 Software Documentation (PN05492D DCN6485) A-33 Name Description Units XIN76 Channel 7 Analog In

Page 251 - 10.2.4. OPTICAL FILTERS

Teledyne API T/E-Series Models 101, 102, and 108 Software Documentation (PN05492D DCN6485) A-34 APPENDIX A-7: MODBUS Register Map MODBUS Register

Page 252 - 10.2.5. OPTICAL LENSES

Teledyne API T/E-Series Models 101, 102, and 108 Software Documentation (PN05492D DCN6485) A-35 58 Sample pressure “Hg 60 Internal box temperature

Page 253

Teledyne API T/E-Series Models 101, 102, and 108 Software Documentation (PN05492D DCN6485) A-36 4 PMT temperature warning 5 Reaction cell temperat

Page 254 - 10.2.6.5. Light Pollution

Teledyne API T/E-Series Models 101, 102, and 108 Software Documentation (PN05492D DCN6485) A-37 1 T101, M101E. 2 Concentration alarm option. 3 Set

Page 255 - 10.3. PNEUMATIC OPERATION

Teledyne API T/E-Series Models 101, 102, and 108 Software Documentation (PN05492D DCN6485) A-38 This page intentionally left blank. 07266B DCN6485

Page 256 - 10.3.1. SAMPLE GAS FLOW

T101 Spare Parts ListPARTNUMBER DESCRIPTION000940100 CD, ORIFICE, .003 GREEN 000940400 CD, ORIFICE, .004 BLUE 000

Page 257 - 10.3.3. FLOW RATE CONTROL

Teledyne API – T101 Operation Manual Getting Started 31 Table 3-2. Rear Panel Description Component Function cooling fan Pulls ambient air into cha

Page 258

T101 Spare Parts ListPARTNUMBER DESCRIPTION(Reference: 073470000, 1/19/2011 10:10:13 AM) 058021100 PCA, MOTHERBD, GEN 5-ICOP

Page 259 - 10.3.6. SO

T101 Spare Parts ListPARTNUMBER DESCRIPTION(Reference: 073470000, 1/19/2011 10:10:13 AM) OR0000094 ORING, 2-228V, 50 DURO VITON(KB)

Page 260 - 10.3.7. PNEUMATIC SENSORS

This page intentionally left blank. B-407266B DCN6845

Page 261 - CPU Card

Appendix C Warranty/Repair Questionnaire T101, M101E (05494D DCN5798) C-1 Company: _________________________ Contact Name: ______________________

Page 262

Appendix C Warranty/Repair Questionnaire T101, M101E (05494D DCN5798) C-2 Notes and further information: _________________________________________

Page 263 - 10.4.1. CPU

APPENDIX D – Wire List and Electronic Schematics 07266B DCN6845D-1

Page 264

This page intentionally left blank. D-207266B DCN6845

Page 265 - Figure 10-13. PMT Assembly

Interconnect List, T101(Reference 0736501 Rev01)RevisionChecked DateDCN01 KV 9/3/10 Cable PN Signal AssemblyPN J/P PinAssemblyPN J/P Pin0337801 CBL A

Page 266 - 10.4.5. PMT COOLING SYSTEM

Interconnect List, T101(Reference 0736501 Rev01)Cable PN Signal AssemblyPN J/P PinAssemblyPN J/P PinFROMTO04176CBL, DC POWER TO RELAY BOARDDGND Relay

Page 267 - 10.4.6. PMT PREAMPLIFIER

Interconnect List, T101(Reference 0736501 Rev01)Cable PN Signal AssemblyPN J/P PinAssemblyPN J/P PinFROMTO04562 CBL, Z/S IZS VALVESSample Valve +12VRe

Page 268 - PMT Preamp PCA

Getting Started Teledyne API – T101 Operation Manual 32 3.2.3. INTERNAL CHASSIS LAYOUT Figure 3-5. Internal Chassis Layout 07266B DCN6485

Page 269 - 10.4.8.2. Valve Control

Interconnect List, T101(Reference 0736501 Rev01)Cable PN Signal AssemblyPN J/P PinAssemblyPN J/P PinFROMTO06746CBL, MB TO 06154 CPUGND Motherboard 058

Page 270

07266B DCN6845D-7

Page 271 - 10.4.10.2. Sensor Inputs

1 2 3 4 5 6ABCD654321DCBATitleNumber RevisionSizeBDate: 30-Jun-2004 Sheet of File: N:\PCBMGR\RELEASED\03954cc\PROTEL\03954a.ddbDrawn By:TeTe03956 A

Page 272 - 10.4.11. ANALOG OUTPUTS

1 2 3 4 5 6ABCD654321DCBATitleNumber RevisionSizeBDate: 30-Jun-2004 Sheet of File: N:\PCBMGR\RELEASED\03954cc\PROTEL\03954a.ddbDrawn By:TeTe03956 A

Page 273 - 10.4.14. POWER UP CIRCUIT

1 2 3 4 5 6ABCD654321DCBATitleNumber RevisionSizeBDate: 30-Jun-2004 Sheet of File: N:\PCBMGR\RELEASED\03954cc\PROTEL\03954a.ddbDrawn By:TeTe03956 A

Page 274

1 2 34ABCD4321DCBAAPPROVALSDRAWNCHECKEDAPPROVEDDATESIZE DRAW ING NO. REVISIONSHEETThe information herein is theproperty of API and issubmitted in stri

Page 275

1 2 3 4 5 6ABCD654321DCBATitleNumber RevisionSizeBDate: 10-May-2007 Sheet of File: N:\PCBMGR\04179cc\Source\RevG\04179.ddbDrawn By:H04181M100E/200E

Page 276 - 10.7.1. ADAPTIVE FILTER

1 2 3 4 5 6ABCD654321DCBATitleNumber RevisionSizeBDate: 10-May-2007 Sheet of File: N:\PCBMGR\04179cc\Source\RevG\04179.ddbDrawn By:R27 499R8150KR15

Page 277

1 2 3 4 5 6ABCD654321DCBATitleNumber RevisionSizeBDate: 10-May-2007 Sheet of File: N:\PCBMGR\04179cc\Source\RevG\04179.ddbDrawn By:R121000MR34.99KR

Page 278 - FEATURE

1 2 3 4 5 6ABCD654321DCBATitleNumber RevisionSizeBDate: 10-May-2007 Sheet of File: N:\PCBMGR\04179cc\Source\RevG\04179.ddbDrawn By:48231U16ALF353,

Page 279

Teledyne API – T101 Operation Manual Getting Started 33 3.3. ELECTRICAL CONNECTIONS NOTE To maintain compliance with EMC standards, it is required

Page 280

1 2 3 4ABCD4321DCBAAPPROVALSDRAWNCHECKEDAPPROVEDDATESIZE DRAWING NO. REVISIONSHEETThe information herein is theproperty of API and issubmitted in stri

Page 281

1 2 3 4 5 6ABCD654321DCBAAPPROVALSDRAWNCHECKEDAPPROVEDDATESIZE DRAWING NO. REVISIONSHEETError : LOGO.BMP file not found.The information herein is thep

Page 282 - 11.4.1. GENERAL RULES

1 2 3 4 5 6ABCD654321DCBATitleNumber RevisionSizeBDate: 2-Jun-2004 Sheet of File: N:\PCBMGR\RELEASED\04691CC\source\04693e.ddbDrawn By:VIN5GND3SW4

Page 283

12345 6ABCD654321DCBATitleNumber RevisionSizeBDate: 13-Jan-2005 Sheet of File: N:\PCBMGR\UNREL\04930PW\Protel\04930.DDBDrawn By:12J2+15TP2TP3C160.1

Page 284 - AND MAINTENANCE

11223344D DC CB BA ATi tleNumber Revis ionSizeADate: 6/28/2004 Sheet of File: N:\PCBMGR\..\04468B.sch Drawn By:12345678JP1No t U sedR122R2SCH, E-S

Page 285

11223344D DC CB BA ATitleNumber RevisionSizeADate: 1/28/2010 Sheet ofFil e: N:\PCBMGR\..\06731-1_ETHERNET.SchDocDrawn By:Auxiliary I/O Board

Page 286 - DO NOT use pink-poly bags

11223344D DC CB BA ATitleNum ber RevisionSizeADate: 1/28/2010 Sh eet ofFil e: N:\PCBMGR\..\06731-2_USB.SchDoc Drawn By:Auxiliary I/O Board

Page 287 - GLOSSARY

11223344D DC CB BA ATitleNumber RevisionSizeADate: 1/28/2010 Sheet ofFi le: N:\PCBM G R \..\06731-3_ ADC.Sch Doc Drawn By:Auxiliary I/O Boar

Page 288

112233445566D DC CB BA ATitleNumber RevisionSizeBDate: 6/24/2010 Sheet ofFile: N:\PCBMGR\..\06696.P1.R3.schdoc Drawn By:GUI Interface06698D14

Page 289

112233445566D DC CB BA ATitleNumber RevisionSizeBDate: 6/24/2010 Sheet ofFile: N:\PCBMGR\ ..\06696.P2.R3.schdoc D rawn By:06698D2422uHL1Vin5SH

Page 290

Getting Started Teledyne API – T101 Operation Manual 34 Table 3-3. Analog Input Pin Assignments PIN DESCRIPTION DAS PARAMETER1 1 Analog input # 1

Page 291

112233445566D DC CB BA ATitleNumber RevisionSizeBDate: 6/24/2010 Sheet ofFile: N:\PCBMGR\..\06696.P3.R3.schdoc Drawn By:06698D4312345To old

Page 292

112233445566D DC CB BA ATitleNumber RevisionSizeBDate: 6/24/2010 Sheet ofFile: N:\PCBMGR\ ..\06696.P4.R3.schdoc D rawn By:06698D44TOUCH SCREE

Page 293

    $%&''&%$7LWOH1XPEHU 5HYLVLRQ6L]H2UFDG%'DWH 0D\ 6KHHWRI)LOH 1?3&%0*5?5-(VHULHV0RWKH

Page 294 - (Primary Setup Menu)

    $%&''&%$7LWOH1XPEHU 5HYLVLRQ6L]H2UFDG%'DWH 0D\ 6KHHWRI)LOH 1?3&%0*5?5-(VHULHV0RWKH

Page 295

    $%&''&%$7LWOH1XPEHU 5HYLVLRQ6L]H2UFDG%'DWH 0D\ 6KHHWRI)LOH 1?3&%0*5?5-(VHULHV0RWKH

Page 296

   > $$>+?@ !FGJMQ XZX[ &?Q Q\  !\ !8 X X G

Page 297 - ENTER SETUP PASS: 818

    $%&''&%$7LWOH1XPEHU 5HYLVLRQ6L]H2UFDG%'DWH 0D\ 6KHHWRI)LOH 1?3&%0*5?5-(VHULHV0RWKH

Page 298

    $%&''&%$7LWOH1XPEHU 5HYLVLRQ6L]H2UFDG%'DWH 0D\ 6KHHWRI)LOH 1?3&%0*5?5-(VHULHV0RWKH

Page 299 - REPORTED

    $%&''&%$7LWOH1XPEHU 5HYLVLRQ6L]H2UFDG%'DWH 0D\ 6KHHWRI)LOH 1?3&%0*5?5-(VHULHV0RWKH

Page 300

    $%&''&%$7LWOH1XPEHU 5HYLVLRQ6L]H2UFDG%'DWH 0D\ 6KHHWRI)LOH 1?3&%0*5?5-(VHULHV0RWKH

Page 301

Teledyne API – T101 Operation Manual Getting Started 35 Table 3-4. Analog Output Pin Assignmentss PIN ANALOG OUTPUT VOLTAGE OUTPUT CURRENT LOOP

Page 302

    $%&''&%$7LWOH1XPEHU 5HYLVLRQ6L]H2UFDG%'DWH 0D\ 6KHHWRI)LOH 1?3&%0*5?5-(VHULHV0RWKH

Page 303

11223344D DC CB BA ATitleNumber RevisionSizeADate: 5/7/2010 Sheet ofFil e: N:\PCBMGR\..\06882-P1-R0.SchDoc Drawn By:LVDS, Transmitter BoardB

Page 304

Getting Started Teledyne API – T101 Operation Manual 36 3.3.3. CONNECTING THE STATUS OUTPUTS The analyzer’s status outputs are accessed through a

Page 305

Teledyne API – T101 Operation Manual Getting Started 37 Table 3-5. Status Output Signals REAR PANEL LABEL STATUS DEFINITION CONDITION 1 SYSTEM OK

Page 306

ii This page intentionally left blank. 07266B DCN6485

Page 307

Getting Started Teledyne API – T101 Operation Manual 38 SPAN CALZERO CALSPAN CALZERO CALCONTROL IN A B C D E F U

Page 308

Teledyne API – T101 Operation Manual Getting Started 39 3.3.5. CONNECTING THE COMMUNICATIONS PORTS 3.3.5.1. Connecting the Serial Ports To utilize

Page 309

Getting Started Teledyne API – T101 Operation Manual 40 Table 3-7. Inlet / Outlet Connector Descriptions REAR PANEL LABEL FUNCTION SAMPLE Co

Page 310

Teledyne API – T101 Operation Manual Getting Started 41 ValveNeedle valve to control flow Chassis SampleZero Air Ex haustSpanMODEL 701 Zero Air Ge

Page 311

Getting Started Teledyne API – T101 Operation Manual 42 PMT EXHA UST THROUGH OUTERLAYER OF KICK ER REACTION CELL PURGE FL O W SENSORFLOW /

Page 312

Teledyne API – T101 Operation Manual Getting Started 43 CAUTION The exhaust from the external pump needs to be vented outside the immediate area o

Page 313

Getting Started Teledyne API – T101 Operation Manual 44 Table 3-9. NIST-SRM's Available for Traceability of H2S & SO2 Calibration Gases N

Page 314

Teledyne API – T101 Operation Manual Getting Started 45 3.4.1.1. Connections with Internal Valve Options Installed This section covers pneumatic co

Page 315

Getting Started Teledyne API – T101 Operation Manual 46 ZERO/SPAN (Z/S) VALVE GAS FLOW PMT EXHAUST TO OUTER LAYER OF KICKER REACTION CELL PURG

Page 316

Teledyne API – T101 Operation Manual Getting Started 47 The state of the zero/span valves can also be controlled:  Manually from the analyzer’s f

Page 317

iii SAFETY MESSAGES Important safety messages are provided throughout this manual for the purpose of avoiding personal injury or instrument damage.

Page 318

Getting Started Teledyne API – T101 Operation Manual 48 Temperature of the H2S: Increasing the temperature of the increases the pressure inside th

Page 319

Teledyne API – T101 Operation Manual Getting Started 49 PMT EXHAUST TO OUTER LAYER OF KIC KER REA CTION CELL PUR GE FLOW SENSORFLOW / PRESSURE

Page 320 - Name Description

Getting Started Teledyne API – T101 Operation Manual 50 The state of the IZS valves can also be controlled:  Manually from the analyzer’s front

Page 321

Teledyne API – T101 Operation Manual Getting Started 51 Table 3-12. Possible Warning Messages at Start-Up WARNING MESSAGE MEANING ANALOG CAL WARNI

Page 322 - Name Description Units

Getting Started Teledyne API – T101 Operation Manual 52 3.5.4. FUNCTIONAL CHECK 1. After the analyzer’s components have warmed up for at least 30

Page 323

Teledyne API – T101 Operation Manual Getting Started 53 3.6. INITIAL CALIBRATION 3.6.1. BASIC CALIBRATION PROCEDURE The following three-step proced

Page 324

Getting Started Teledyne API – T101 Operation Manual 54 STEP TWO: Set the expected H2S span gas concentration. This sequence causes the analyzer

Page 325

Teledyne API – T101 Operation Manual Getting Started 55 STEP THREE: Perform the zero/span calibration procedure: M-P CAL H2S STB=X

Page 326 - (single-bit; read/write)

Getting Started Teledyne API – T101 Operation Manual 56 3.6.2. INTERFERENCES FOR H2S MEASUREMENTS It should be noted that the fluorescence method

Page 327

57 4. OPERATING INSTRUCTIONS The T101 is a micro-computer-controlled analyzer with a dynamic menu interface for easy and yet powerful and flexible

Page 328

iv CONSIGNES DE SÉCURITÉ Des consignes de sécurité importantes sont fournies tout au long du présent manuel dans le but d’éviter des blessures cor

Page 329 - T101 Spare Parts List

Operating Instructions Teledyne API – T101 Operation Manual 58 MODE DESCRIPTION SETUP X.#2 SETUP mode is being used to configure the analyzer. Th

Page 330

Teledyne API – T101 Operation Manual Operating Instructions 59 Table 4-2. Test Functions Defined DISPLAY PARAMETER UNITS DESCRIPTION RANGE Range

Page 331

Operating Instructions Teledyne API – T101 Operation Manual 60 TEMP Converter Temperature H2S present in the sample gas into SO2. TEST2 Test Sign

Page 332 - 07266B DCN6845

Teledyne API – T101 Operation Manual Operating Instructions 61 4.2.2. WARNING MESSAGES The most common instrument failures will be reported as a war

Page 333 - T101, M101E

Operating Instructions Teledyne API – T101 Operation Manual 62 To view and clear warning messages: SAMPLE HVPS WARNING

Page 334

Teledyne API – T101 Operation Manual Operating Instructions 63 The default status of the calibration password is OFF. To enable the calibration pas

Page 335

Operating Instructions Teledyne API – T101 Operation Manual 64 4.4. SETUP MODE The SETUP mode allows you to configure the analyzer’s hardware and

Page 336

Teledyne API – T101 Operation Manual Operating Instructions 65 NOTE If the analyzer beeps when you press the EXIT button, it means that you’ve made

Page 337

Operating Instructions Teledyne API – T101 Operation Manual 66 4.4.2. SETUP – ACAL: AUTO CALIBRATION Used to set up and operate the internal valve

Page 338

Teledyne API – T101 Operation Manual Operating Instructions 67 AUTO mode, see section 4.4.4.6) or operate independently (IND mode, see Section 4.4.

Page 339

v WARRANTY WARRANTY POLICY (02024 F) Teledyne Advanced Pollution Instrumentation (TAPI), a business unit of Teledyne Instruments, Inc., provides th

Page 340

Operating Instructions Teledyne API – T101 Operation Manual 68 4.4.4.3. Reporting Range Modes The T101 provides three analog output range modes to

Page 341

Teledyne API – T101 Operation Manual Operating Instructions 69 4.4.4.4. Single Range Mode (SNGL) The default range mode for the analyzer is single r

Page 342

Operating Instructions Teledyne API – T101 Operation Manual 70 4.4.4.5. Independent Range Mode (IND) Selecting independent range mode allows the A

Page 343

Teledyne API – T101 Operation Manual Operating Instructions 71 NOTE On analyzers with the multigas option activated (see Sections 4.5.1 and 6.8) th

Page 344

Operating Instructions Teledyne API – T101 Operation Manual 72 NOTE On analyzers with the multigas option activated (see Section s 4.5.1 and 6.8)

Page 345

Teledyne API – T101 Operation Manual Operating Instructions 73 4.4.4.8. Dilution Ratio The dilution ratio is a software option that allows the user

Page 346

Operating Instructions Teledyne API – T101 Operation Manual 74 4.4.5. SETUP – PASS: PASSWORD PROTECTION The menu system provides password protecti

Page 347

Teledyne API – T101 Operation Manual Operating Instructions 75 4.4.6. SETUP – CLK: SETTING THE INTERNAL TIME-OF-DAY CLOCK The T101 has a built-in cl

Page 348

Operating Instructions Teledyne API – T101 Operation Manual 76 In order to compensate for CPU clocks which run fast or slow, there is a variable t

Page 349

Teledyne API – T101 Operation Manual Operating Instructions 77 4.5. SETUP – VARS: USING THE INTERNAL VARIABLES The T101 has several-user adjustable

Page 350

vi This page intentionally left blank. 07266B DCN6485

Page 351

Operating Instructions Teledyne API – T101 Operation Manual 78 NO. VARIABLE DESCRIPTION ALLOWED VALUES 11 TIME_SINCE_SVC Displays time in hours s

Page 352

Teledyne API – T101 Operation Manual Operating Instructions 79 To access and navigate the VARS menu, use the following touchscreen button sequence:

Page 353

Operating Instructions Teledyne API – T101 Operation Manual 80 4.5.1. SETTING THE GAS MEASUREMENT MODE If the switching valves software is activat

Page 354

Teledyne API – T101 Operation Manual Operating Instructions 81 4.6. SETUP – DIAG: USING THE DIAGNOSTICS FUNCTIONS A series of diagnostic tools is g

Page 355

Operating Instructions Teledyne API – T101 Operation Manual 82 To access the DIAG functions press the following buttons: SAMPLE ENTE

Page 356 - Auxiliary I/O Board (USB)

Teledyne API – T101 Operation Manual Operating Instructions 83 4.6.1. SIGNAL I/O The signal I/O diagnostic mode allows reviewing and changing the di

Page 357 - Auxiliary I/O Board (ADC)

Operating Instructions Teledyne API – T101 Operation Manual 84 4.6.2. ANALOG OUTPUT STEP TEST This test can be used to check the accuracy and prop

Page 358 - GUI Interface

Teledyne API – T101 Operation Manual Operating Instructions 85 4.6.3. ANALOG I/O CONFIGURATION Table 4-8 lists the analog I/O functions that are ava

Page 359

Operating Instructions Teledyne API – T101 Operation Manual 86 The following DC current output limits apply to the current loop modules: Table 4-1

Page 360 - MBUS B B

Teledyne API – T101 Operation Manual Operating Instructions 87 4.6.3.1. Analog Output Signal Type and Range Span Selection To select an output signa

Page 361

vii ABOUT THIS MANUAL This T101 operation manual, PN 07266, is comprised of multiple documents in PDF format, as listed below. Part No. Rev Name/

Page 362

Operating Instructions Teledyne API – T101 Operation Manual 88 4.6.3.2. Analog Output Calibration Mode Analog output calibration should be carried

Page 363

Teledyne API – T101 Operation Manual Operating Instructions 89 To automatically calibrate an single analog channel, activate the ANALOG I/O CONFIGUR

Page 364

Operating Instructions Teledyne API – T101 Operation Manual 90 To select manual output calibration for a particular channel, activate the ANALOG I

Page 365

Teledyne API – T101 Operation Manual Operating Instructions 91 V +DC Gnd V OUT + V OUT -V IN + V IN - Recording Device ANALYZERSee Table 3-1

Page 366

Operating Instructions Teledyne API – T101 Operation Manual 92 4.6.3.4. Analog Output Offset Adjustment Some analog signal recorders require that

Page 367

Teledyne API – T101 Operation Manual Operating Instructions 93 Similar to the voltage calibration, the software allows this current adjustment to be

Page 368

Operating Instructions Teledyne API – T101 Operation Manual 94 To adjust the zero and span values of the current outputs, activate the ANALOG I/O

Page 369

Teledyne API – T101 Operation Manual Operating Instructions 95 4.6.3.6. AIN Calibration This is the sub-menu to conduct the analog input calibration

Page 370

Operating Instructions Teledyne API – T101 Operation Manual 96 4.6.3.7. Analog Inputs (XIN1…XIN8) Option Configuration To configure the analyzer’s

Page 371 - To LCD Display

Teledyne API – T101 Operation Manual Operating Instructions 97 4.6.4. OPTIC TEST The optic test function tests the response of the PMT sensor by tur

Commentaires sur ces manuels

Pas de commentaire