Calibration Procedures Teledyne API T802 Paramagnetic O
2
Analyzer Operation Manual
180 07275A (DCN6005)
9.5.3. ACTIVATE CO
2
SENSOR STABILITY FUNCTION
To change the stability test function from O
2
concentration to the CO
2
sensor
output, press:
Press 3
times to return
to
menu
SETUP X.X
CO2 O2 ENTR
<TST TST> SETUP
SETUP X.X
CFG DAS ACAL RANG PASS CLK EXIT
SETUP X.X
COMM DIAG ALRM EXIT
SETUP X.X 0) DAS_HOLD_OFF=15.0 Minutes
<PREV
JUMP EDIT PRNT EXIT
SETUP X.X
<PREV NEXT> JUMP PRNT EXIT
SETUP X.X
O2 ENTR EXIT
SETUP X.X
EXIT
CO
2
and O
2
options only
appear if
associated
sensors are
installed.
NOTE
Use the same procedure to reset the STB test function to O
2
when the CO
2
calibration procedure is
complete.
07275B DCN6418