Configurable Analog Output Overview Teledyne API Model M201E NH
3
Analyzer Operator Manual
26 05206H DCN5910
Table 3-2. Analog Output Pin-Outs
PIN ANALOG OUTPUT VOLTAGE SIGNAL CURRENT SIGNAL
1 V Out I Out +
2
A1
Ground I Out -
3 V Out I Out +
4
A2
Ground I Out -
5 V Out I Out +
6
A3
Ground I Out -
7 V Out N/A
8
A4
Ground N/A
Additionally A1, A2 and A3 may be equipped with optional 0-20 mA current loop
drivers. The 4-20 mA option is not available on A4.
Note: In actuality the analog output configuration of the analyzer may be different
than stated above. The outputs can be configured differently at the factory
depending on whether they were assigned during the procurement of the product.
It is possible to check the configuration of the analyzer by accessing the ANALOG
I/O CONFIGURATION through the DIAG menu.
The analyzer operates in Dual Mode during gas detection. This means each of the
measured concentrations can have two separate slopes and offsets, one for the
low range and one for the high range. Though uncommon, a user may decide to
calibrate the analyzer with nitric oxide at, for example, 100 PPB using the LOW
range and then perform another calibration at 450 PPB using the HIGH range.
Through the analog outputs the user can then assign analog output A1 to
TN
x
CNC1 and output A2 to TNxCNC2. Gases with the “1” designation use the
slope and offset for the LOW range, while gases with the “2” designation will use
the slope and offset for the HIGH range
.
It is recommended that both the LOW and
HIGH ranges be calibrated at the same time. Independent of whether the HIGH
range is actually being outputted to the analog outputs.