Teledyne API Model T360/T360M Operation Manual Principles of Operation
203
7.3.6. Pneumatic Operation of the O
2
Sensor
Pneumatically, the O
2
sensor is connected to the bypass manifold and draws a
flow of about 120 cm³/min in addition to the normal sample flow rate and is
separately controlled with its own critical flow orifice. Figure 7-10 shows the
internal pneumatics of the analy
zer with the O
2
sensor installed.
Figure 7-10: Gas Flow with O2 Sensor Option
07272B DCN6552