Teledyne API Model T360/T360M Operation Manual Getting Started
35
3.5. ELECTRICAL CONNECTIONS
WARNING
Never disconnect PCAs, wiring harnesses or electronic subassemblies
while the instrument is under power.
NOTE
To maintain compliance with EMC standards, it is required that the cable length be no greater than 3
meters for all I/O connections, which include Analog In, Analog Out, Status Out, Control In, Ethernet/LAN,
USB, RS-232, and RS-485.
Refer to Figure 3-4 for the locations of the rear panel connections.
3.5.1. Power Connection
The T360 analyzer can be configured for both 100-130 V and 210-240 V at either
50 or 60 Hz. To avoid damage to your analyzer, make sure that the AC power
voltage matches the voltage indicated on the rear panel serial number label and
that the frequency is between 47 and 63 Hz.
Attach the power cord to the analyzer and plug it into a power outlet capable of
carrying at least 10 A current at your AC voltage and that it is equipped with a
functioning earth ground.
WARNING
Power connection must have functioning ground connection.
Do not defeat the ground wire on power plug.
Turn off analyzer power before disconnecting or
connecting electrical subassemblies.
Do not operate with cover off.
CAUTION
Check the voltage and frequency specifications on the rear panel Model
Label for compatibility with the local power before plugging the analyzer
into line power.
Do not plug in the power cord if the voltage or frequency is incorrect.
07272B DCN6552