Teledyne API Model T360/T360M Operation Manual Operating Instructions
93
4.7.2.13. HOLDOFF FEATURE
The DAS HOLDOFF feature allows to prevent data collection during calibrations
and during the DAS_HOLDOFF period enabled and specified in the VARS
(Section 4.12). To enable or disable the HOLDOFF, follow the instruction shown
in Section 4.7.2.2 then press:
Edit Data Channel Menu
SETUP X.X 0) CONC: ATIMER, 1, 900
PREV NEXT INS DEL EDIT PRNT EXIT
Exits to the main Data
Acquisition menu
SETUP X.X NAME:CONC
<SET SET> EDIT PRINT EXIT
SETUP X.X CAL HOLD OFF:ON
SET> EDIT PR INT EXIT
ENTR accepts the new
setting and returns to the
previous me nu.
EXIT ignores the new setting
and returns to the previous
menu.
SETUP X.X CAL HOLD OFF:ON
ON ENTR EXIT
Toggle to turn HOLDOFF
ON or OFF
Press SET> until…
From the DATA ACQUISITION menu
(see Section 4.7.2.2)
07272B DCN6552