Teledyne API Model T360/T360M Operation Manual Getting Started
61
3.8.1. Initial O
2
Sensor Calibration Procedure
If your instrument includes the O
2
sensor option, that should be calibrated as well.
3.8.1.1. O
2
SENSOR CALIBRATION SETUP
The pneumatic connections for calibrating are as follows:
Figure 3-21: O2 Sensor Calibration Set Up
O
2
SENSOR ZERO GAS: Teledyne Instruments’ recommends using pure N
2
when calibration the zero point of your O
2
sensor option.
O
2
SENSOR SPAN GAS: Teledyne Instruments’ recommends using 21% O
2
in
N
2
when calibration the span point of your O
2
sensor option.
07272B DCN6552