Teledyne T101 Manuel d'utilisateur Page 85

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 371
  • Table des matières
  • DEPANNAGE
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 84
Teledyne API – T101 Operation Manual Operating Instructions
83
4.6.1. SIGNAL I/O
The signal I/O diagnostic mode allows reviewing and changing the digital and analog
input/output functions of the analyzer. See Appendix A for a list of the parameters
available under this menu.
NOTE
Any changes of signal I/O settings will remain in effect only until the signal I/O menu is exited. Exceptions
are the ozone generator override and the flow sensor calibration, which remain as entered when exiting.
To enter the signal I/O test mode, press:
SAMPLE RANGE = 500.0 PPB H2S =XXX.X
< TST TST > CAL SETUP
EXAMPLE
SAMPLE ENTER SETUP PASS : 818
8 1 8 ENTR EXIT
SETUP X.X PRIMARY SETUP MENU
CFG DAS RNGE PASS CLK MORE EXIT
DIAG SIGNAL I / O
PREV NEXT JUMP ENTR EXIT
DIAG I / O EXT_ZERO_CAL=OFF
PREV NEXT JUMP PRNT EXIT
Press JUMP to go
dir ectly to a specific
signal
See Appendix A-4 for
a complete li st of
ava ilable SIGNALS
DIAG I / O JUMP TO: 12
1 2 ENTR EXIT
EXAMPLE :
Enter 12 to Jump to
12) ST_CONC_VALID
DIAG I / O ST_CONC_VALID = ON
PREV NE XT JUMP ON PR NT EXIT
Exit to return
to t h e
DIAG menu
SETUP X.X SECONDARY SETUP MENU
COMM VARS DIAG EXIT
Pr ess NEXT & PREV to
m ove between sign al
types.
Pressing PRNT will send a formatted printout to the serial port and can be captured
wi th a comp uter or other output device .
EXIT returns
to the main
SAMPLE display
07266B DCN6485
Vue de la page 84
1 2 ... 80 81 82 83 84 85 86 87 88 89 90 ... 370 371

Commentaires sur ces manuels

Pas de commentaire