Calibration Procedures Teledyne API – T101 Operation Manual
166
STEP ONE: Connect the sources of zero air and span gas as shown below.
No Valve Options Installed
Chassis
Sample
Zero Air
Exhaust
S
p
an
MODEL 701
Zero Air
Generator
Calibrated
H
2S GAS
(At high
concentration)
Source of
SAMPLE Gas
Removed
during
Calibration
Val ve
OR
Needle valve to
control flow
Chassis
Sam
p
le
Zero Air
Exhaust
Span
MODEL 701
Zero Air
Generator
VENT
Calibrated
H
2
S GAS
(At span gas
concentr ation)
Source of
SAMPLE Gas
Removed
during
ca libration
MO DEL T700 Gas
Dilution
Calibrator
Figure 6-1. Setup for Manual Calibration without Z/S Valve or IZS Option
07266B DCN6485